Lonizing radiation measurement solution in a hospital environment
dc.contributor.author | García Sánchez, Antonio Javier | |
dc.contributor.author | García Angosto, Enrique Ángel | |
dc.contributor.author | Moreno Riquelme, Pedro Antonio | |
dc.contributor.author | Serna Berna, Alfredo | |
dc.contributor.author | Ramos Amores, David | |
dc.date.accessioned | 2020-07-15T08:54:56Z | |
dc.date.available | 2020-07-15T08:54:56Z | |
dc.date.issued | 2018-02 | |
dc.identifier.citation | Garcia-Sanchez, A.-J.; Garcia Angosto, E.A.; Moreno Riquelme, P.A.; Serna Berna, A.; Ramos-Amores, D. Ionizing Radiation Measurement Solution in a Hospital Environment. Sensors 2018, 18, 510. | es_ES |
dc.identifier.issn | 1424-8220 | |
dc.description.abstract | Ionizing radiation is one of the main risks affecting healthcare workers and patients worldwide. Special attention has to be paid to medical staff in the vicinity of radiological equipment or patients undergoing radioisotope procedures. To measure radiation values, traditional area meters are strategically placed in hospitals and personal dosimeters are worn by workers. However, important drawbacks inherent to these systems in terms of cost, detection precision, real time data processing, flexibility, and so on, have been detected and carefully detailed. To overcome these inconveniences, a low cost, open-source, portable radiation measurement system is proposed. The goal is to deploy devices integrating a commercial Geiger-Muller (GM) detector to capture radiation doses in real time and to wirelessly dispatch them to a remote database where the radiation values are stored. Medical staff will be able to check the accumulated doses first hand, as well as other statistics related to radiation by means of a smartphone application. Finally, the device is certified by an accredited calibration center, to later validate the entire system in a hospital environment | es_ES |
dc.description.sponsorship | This research has been supported by the project AIM, ref. TEC2016-76465-C2-1-R (AEI/FEDER, UE). This work is dedicated to the memory of Felipe Garcia-Sanchez. | es_ES |
dc.format | application/pdf | es_ES |
dc.language.iso | eng | es_ES |
dc.relation.ispartof | Contribución a la dosimetría personal en pruebas médicas radiológicas para entornos regionales | es_ES |
dc.relation.uri | http://hdl.handle.net/10317/8631 | es_ES |
dc.rights | Atribución-NoComercial-SinDerivadas 3.0 España | * |
dc.rights.uri | http://creativecommons.org/licenses/by-nc-nd/3.0/es/ | * |
dc.title | Lonizing radiation measurement solution in a hospital environment | es_ES |
dc.type | info:eu-repo/semantics/article | es_ES |
dc.subject.other | Ingeniería Telemática | es_ES |
dc.subject | Gamma radiation | es_ES |
dc.subject | Healthcare workers safety | es_ES |
dc.subject | Dosimetry solution | es_ES |
dc.subject | Verification | es_ES |
dc.subject | Evaluation | es_ES |
dc.identifier.uri | http://hdl.handle.net/10317/8654 | |
dc.identifier.doi | 10.3390/s18020510 | |
dc.identifier.url | https://www.mdpi.com/1424-8220/18/2/510#cite | |
dc.rights.accessRights | info:eu-repo/semantics/openAccess | es_ES |
dc.type.version | info:eu-repo/semantics/publishedVersion | es_ES |
dc.relation.projectID | TEC2016-76465-C2-1-R | es_ES |
dc.subject.unesco | 3201.11 Radiología | es_ES |
Ficheros en el ítem
Este ítem aparece en la(s) siguiente(s) colección(ones)
-
Artículos [1733]
Redes sociales